F2 3E 1

F2 3E 1
F2 3E 1
Codice prodotto
F2 3E 1
Produttore
City Technology - Cella elettrochimica per Fluoro

Measuring Range: 0-1 ppm
Sensitivity Range: 100 nA/ppm ± 300 nA/ppm (negative current)
Zero Current at 20°C: < ± 20 nA
Resolution at 20°C: < 0.02 ppm
Bias Potential: 0 mV
Linearity: < 5% full scale

Response Time at 20°C:

  • t50 < 30 s calculated from 4 min. exposure time with 1 ppm Cl2
  • t90 < 80 s calculated from 4 min. exposure time with 1 ppm Cl2

Long Term Sensitivity Drift: < 5% month

Operation Conditions
Temperature Range: -10°C to +40°C
Humidity Range: 15–90% r.H., non–condensing

Effect of Humidity: Abrupt changes of rel. humidity will cause short term drift in zero reading

Sensor Life Expectancy: > 18 months
Warranty: 12 months

Disponibile in versione Classica o con filtro

Life expectancy
> 18 mesi
Range di misura
0-1 ppm
Target Gas
Fluoro
Tempo di Risposta
< 80 secondi (T90)
Uscita
1 ± 0.3 μA/ppm
f23e1.pdf203.29 KB
Contattaci

Vuoi una quotazione o hai bisogno di informazioni?

Ho letto e accetto l’informativa relativa al Trattamento dei Dati Personali ai sensi del Regolamento UE 2016/679 artt. 13 e 14 * Informativa privacy