MEASUREMENT
Operating Principle: 3-electrode electrochemical
Nominal Range: 0-1000 ppm NO
Maximum Overload: 5000 ppm NO
Filter: To remove SO2 from gas stream
Sensitivity: 0.10 ± 0.02 μA/ppm
Response Time (T90): <25 Seconds
Baseline Offset (clean air): 0 to 12 ppm equivalent
Zero Shift (+20°C to +40°C): 30 ppm equivalent
Repeatability: 2% of signal
Output Linearity: Linear
ELECTRICAL
Recommended Load Resistor: 10 Ω
Bias Voltage: +300 mV
MECHANICAL
Weight: 22 g
Orientation: Any
ENVIRONMENTAL
Operating Temperature Range: -20°C to +40°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.01% signal / mbar
Operating Humidity Range: 15 - 90% RH non-condensing
LIFETIME
Long Term Sensitivity Drift: <2% signal loss / month
Expected Operating Life: Three years in air
Storage Life: 6 months in CTL container
3NFF
Codice prodotto
3NFF
Produttore
Life expectancy
> 3 anni
Range di misura
0-1000 ppm
Target Gas
Monossido di Azoto
Tempo di Risposta
< 25 secondi (T90)
Uscita
100 ± 20 nA/ppm
Datasheet303.16 KB