Contattaci
2FO
Codice prodotto
2FO
Produttore
Nominal Range: 0-25% Oxygen
Maximum Overload: 30% Oxygen
Expected Operating Life: Two years in air
Output Signal: 0.41 ± 0.07mA in Air
T95 Response Time: <10 seconds (see note)
Temperature Range: -20°C to +45°C
Temperature Coefficient: 0.2% signal/°C
Pressure Range: Atmospheric ± 10%
Pressure Coeffi cient: <0.02% signal/mBar
Operating Humidity: 0 to 99% RH non-condensing
Long Term Output Drift: <5% signal loss/year
Maximum Load Resistor: 100Ω
Storage Life: Six months in CTL container
Recommended Storage Temperature: 0-20°C
Life expectancy
> 24 mesi
Range di misura
0-25% Oxygen
Target Gas
Ossigeno
Tempo di Risposta
< 10 secondi (T90)
Uscita
410 ± 70 μA/ppm