Contattaci
Cl2 3E 10
Codice prodotto
Cl2 3E 10
Produttore
Measuring Range: 0-10 ppm; typically: 0–5 ppm
Sensitivity Range: 450 nA/ppm ± 200 nA/ppm (negative current)
Zero Current at 20°C: < ± 20 nA
Resolution at 20°C: < 0.05 ppm
Bias Potential: 0 mV
Linearity: < 5% full scale
Response Time at 20°C:
- t50 <30 s calculated from 2 min. exposure time
- t90 <60 s calculated from 2 min. exposure time
Long Term Sensitivity Drift: < 10% per 6 months
Operation Conditions
Temperature Range: -20°C to + 40°C
Humidity Range: 15–90% r.H., non–condensing
Effect of Humidity: no effect on zero current
Sensor Life Expectancy: > 24 months in air
Warranty: 12 months
Life expectancy
> 4 anni
Range di misura
0-10 ppm
Target Gas
Cloro
Tempo di Risposta
< 60 secondi (T90)
Uscita
450 ± 200 nA/ppm
Datasheet203.45 KB