MEASUREMENT
Technology: Electrochemical
Measurement Range: 0-25% vol. O2
Maximum Overload: 30% vol. O2
Output Signal: 80 - 130 mA in Air
T90 Response Time: <15 Seconds
T97 Response Time: <35 Seconds
Zero Current (Offset) ( after 3 minute N2): <0.3% vol. O2
Warm-Up Time: Refer to Characterisation Note
Linearity: S = K loge 1/ (1-C)
ELECTRICAL
Power Rating at 20.9% O2: 0.5 mW
Bias Voltage: -600 ± 10 mV
MECHANICAL
Casing Material: ABS/NORYL
Weight: 14 ± 0.2 g
Orientation Sensitivity: <0.2% vol. O2 equivalent
ENVIRONMENTAL
Operating Temperature Range: -40°C to +60°C
Recommended Storage Temp: 0°C to 20°C in original packaging
Thermal Transient (Temp. plunge +22°C to -20°C): <23.5% vol. O2
Operating Pressure Range: Atmospheric ± 20%
Pressure Coefficient: <0.02% signal/mbar
Pressure Transient (60 cm H2O step change): <150% signal change
Operating Humidity Range: 15%rH to 90%rH non-condensing
LIFETIME
Long Term Sensitivity Drift: <5% signal loss over operating life
Expected Operating Life: 7 years in air
5OxLL
Codice prodotto
5OxLL
Produttore
Life expectancy
> 7 anni
Range di misura
0-25% Ossigeno
Target Gas
Ossigeno
Tempo di Risposta
< 15 secondi (T90)
Uscita
80 - 130 μA in air
Datasheet1.27 MB