5NF

5NF
5NF
Codice prodotto
5NF
Produttore
City Technology - Cella elettrochimica per Monossido di Azoto

MEASUREMENT
Operating Principle
: 4-electrode electrochemical
Measurement Range: 0-1000 ppm NO
Maximum Overload: 5000 ppm NO
Filter: To remove effect of SO2
Filter Life: 25,000 ppm hours (1000 ppm SO2 at 200 ml/min)
Sensitivity: 0.10 ± 0.02 μA/ppm
Resolution: 1 ppm
Response Time (T90): < 30 seconds
Baseline Offset (clean air): 0 to 12 ppm equivalent
Maximum Zero Shift (+20°C to +40°C): 30 ppm equivalent
Repeatability: 2% of signal
Linearity: Linear

ELECTRICAL
Recommended Load Resistor:
 10 Ω
Bias Voltage: +300 mV

MECHANICAL
Weight: 13 g nominal
Orientation: Any

ENVIRONMENTAL
Operating Temperature Range: -20°C to +40°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Operating Humidity Range: 15 - 90% RH non-condensing

LIFETIME
Long Term Sensitivity Drift: <2% signal loss/month
Expected Operating Life: Three years in air
Storage Life: 6 months in CTL container

Life expectancy
> 3 anni
Range di misura
0-1000 ppm
Target Gas
Monossido di Azoto
Tempo di Risposta
< 30 secondi (T90)
Uscita
100 ± 20 nA/ppm
Datasheet329.87 KB
Contattaci

Vuoi una quotazione o hai bisogno di informazioni?

Ho letto e accetto l’informativa relativa al Trattamento dei Dati Personali ai sensi del Regolamento UE 2016/679 artt. 13 e 14 * Informativa privacy