Contact us
PH3 3E 5 LT
Product code
PH3 3E 5 LT
Manufacturer
Measuring Range: 0-5 ppm
Sensitivity Range: 2200 nA/ppm ± 500 nA/ppm
Baseline Offset at 20°C: < ± 20 nA
Bias Potential: 0 mV
Linearity: < 10% full scale
Response Time at 20°C: < 30 seconds (t90)
Long Term Sensitivity Drift: < 5% per 6 months
Operation Conditions
Temperature Range: -20°C to +50°C
Humidity Range: 10–95% r.H., non–condensing
Sensor Life Expectancy: 24 months
Warranty: 12 months
Life expectancy
> 24 months
Output
2.2 ± 0.5 μA/ppm
Measurement range
0-5 ppm
Target Gas
Phosphine
Response Time
< 30 seconds (T90)
ph33e5lt.pdf1.23 MB