Contact us
NO2 3E 50
Product code
NO2 3E 50
Manufacturer
Measuring Range: 0–50 ppm
Sensitivity Range: 200 nA/ppm ± 40 nA/ppm (negative signal)
Zero Current at 20°C: < ± 20 nA
Resolution at 20°C: < 0.1 ppm
Bias Potential: 0 mV
Linearity: < 5% full scale
Response Time at 20°C:
- t50 < 10 s calculated from 4 min. exposure time
- t90 < 30 s calculated from 4 min. exposure time
Long Term Sensitivity Drift: < 5% per month
Operation Conditions
Temperature Range: -20°C to +40°C
Humidity Range: 15–90% r.H, non–condensing
Effect of Humidity: an abrupt change of r.H. will cause a short term drift of zero
reading
Sensor Life Expectancy: > 24 months in air
Warranty: 12 months
Life expectancy
> 24 months
Measurement range
0-50 ppm
Target Gas
Nitrogen dioxide
Response Time
< 30 seconds (T90)
Output
200 ± 40 nA/ppm
Datasheet205.55 KB