Contact us
NH3 3E 100 SE
Product code
NH3 3E 100 SE
Manufacturer
Amperometric 3 electrode sensor cell
Measuring Range: 0–100 ppm
Sensitivity Range: 130 nA/ppm ± 30 nA/ppm
Zero Current at 20°C: < ± 100 nA
Resolution at 20°C: <1 ppm
Bias Potential: 0 mV
Linearity: <10% full scale
Response Time at 20°C
t50: <20 s calculated from 5 min. exposure time
t90: <60 s calculated from 5 min. exposure time
Long Term Sensitivity Drift: <5% per 6 months
Temperature Range: -20°C to +40°C
Humidity Range: 15–90% r.H, non–condensing
Effect of Humidity:
an abrupt change of r.H. will cause a short term drift of zero reading
Sensor Life Expectancy: > 24 months in air
Life expectancy
> 24 months
Measurement range
0-100 ppm
Target Gas
Ammonia
Response Time
< 60 seconds (T90)
Output
130 ± 30 nA/ppm
Scheda tecnica208.51 KB