MEASUREMENT
Operating Principle: 3-electrode electrochemical
Measurement Range: 0-100 ppm CO
Maximum Overload: 500 ppm CO
Filter: To remove H2S
Sensitivity: 0.37 ± 0.07 μA/ppm
Resolution: 0.5 ppm (electronics dependant)
Response Time (T90): <50 Seconds
Baseline Offset (clean air): -0.25 to +0.5 ppm equivalent
Zero Shift (+20°C to +40°C): 1 ppm equivalent
Repeatability: 1% of signal
Output Linearity: Linear
ELECTRICAL
Recommended Load Resistor: 10 Ω
Bias Voltage: Not Required
MECHANICAL
Weight: 22 g
Orientation: Any
ENVIRONMENTAL
Operating Temperature Range: -20°C to +50°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.015% signal / mbar
Operating Humidity Range: 15 - 90% RH non-condensing
LIFETIME
Long Term Sensitivity Drift: <2% signal loss/year
Expected Operating Life: Two years in air
Storage Life: 6 months in CTL container
3STF
Product code
3STF
Manufacturer
Life expectancy
> 24 months
Measurement range
0-100 ppm
Target Gas
Sulfur Dioxide
Response Time
< 50 seconds (T90)
Output
370 ± 70 nA/ppm
Datasheet303.64 KB