C/N

C/N
C/N
Product code
C/N
Manufacturer
City Technology - Electrochemical Cell for Oxygen O2

Nominal Range: 0-25% Oxygen
Max Overload: 30% Oxygen
Expected Operating Life: Nine months in Air
Output Signal: 1.10 ± 0.17mA in Air
T95 Response Time: <20 seconds
Temperature Range: -20°C to +50°C
Temperature Coefficient: 0.2% signal/°C
Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.01% signal/mBar
Operating Humidity: 0 to 99% RH non-condensing
Long Term Output Drift: <5% signal loss/year
Maximum Load Resistor: 100Ω
Storage Life: Six months in CTL container
Recommended Storage Temperature: 0-20°C

Life expectancy
> 9 months
Measurement range
0-25% Oxygen
Target Gas
Oxygen
Response Time
< 20 seconds (T95)
Output
1.10 ± 0.17 mA in Air