3ND

3ND
3ND
Product code
3ND
Manufacturer
City Technology - Electrochemical Cell for Nitrogen Dioxide

MEASUREMENT
Operating Principle
: 3-electrode electrochemical
Nominal Range: 0-100 ppm NO2
Maximum Overload: 1000 ppm NO2
Filter: None
Sensitivity: 0.37 ± 0.07 μA/ppm
Response Time (T90): <35 Seconds
Baseline Offset (clean air): 0 to 2 ppm equivalent
Zero Shift (+20°C to +40°C): -1 ppm equivalent 
Resolution: 0.5 ppm (electronics dependant)
Repeatability: 2% of signal
Output Linearity: Linear

ELECTRICAL
Recommended Load Resistor
: 33 Ω
Bias Voltage: Not Required

MECHANICAL
Weight
: 22 g
Orientation: Any

ENVIRONMENTAL
Operating Temperature Range
: -20°C to +50°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.037% signal / mbar
Operating Humidity Range: 15 - 90% RH non-condensing

LIFETIME
Long Term Sensitivity Drift
: <2% signal loss / month
Expected Operating Life: Two years in air
Storage Life: 6 months in CTL container

Life expectancy
> 24 months
Measurement range
0-100 ppm
Target Gas
Nitrogen dioxide
Response Time
< 35 seconds (T90)
Output
370 ± 70 nA/ppm
Datasheet301.44 KB
Contact us

Vuoi una quotazione o hai bisogno di informazioni?

I have read and accepted the information relating to the Processing of Personal Data pursuant to EU Regulation 2016/679 articles 13 and 14 * Privacy information