Contact us
GS+7NH3-1000
Product code
GS+7NH3-1000
Manufacturer
Range 0 – 1000 ppm
Typical Baseline Range (pure air) ±10 ppm NH3 equivalent
Output Signal 8 ± 4 nA/ppm
Linearity Linear
Response Time, T90 <60 s
Maximum Overload 1500 ppm
Long-term Output Drift <20% per annum
Recommended Load Resistor 10 ohms
Repeatability ±10%
Life expectancy
> 24 months
Measurement range
0-1000 ppm
Target Gas
Ammonia
Response Time
< 60 seconds (T90)
Output
8 ± 4 nA/ppm
Datasheet1.3 MB
GS+7NH3-1000 Recommended Circuit21.07 KB