4ND

4ND
4ND
Product code
4ND
Manufacturer
City Technology - Electrochemical Cell for Nitrogen Dioxide NO2

Nominal Range: 0-20 ppm
Maximum Overload: 150 ppm
Expected Operating Life: Two years in air
Output Signal: 0.6 ± 0.15 μA/ppm
Resolution: 0.1 ppm
Temperature Range: -20°C to +50°C
Pressure Range: Atmospheric ± 10%
T90 Response Time: <25 seconds
Relative Humidity Range: 15 to 90% non-condensing
Typical Baseline Range (pure air): -0.2 to +0.2 ppm equivalent
Maximum Zero Shift (+20°C to +40°C): 0.2 ppm equivalent
Long Term Output Drift: <2% signal loss/month
Recommended Load Resistor: 33 W
Bias Voltage: Not required
Repeatability: <2% of signal
Output Linearity: Linear

Life expectancy
> 24 months
Measurement range
0 - 20 ppm
Target Gas
Nitrogen dioxide
Response Time
< 25 seconds (T90)
Output
600 ± 150 nA/ppm