5MF

5MF
5MF
Product code
5MF
Manufacturer
City Technology - Electrochemical Cell for Carbon Monoxide

PERFORMANCE CHARACTERISTICS
Nominal Range: 
0-40,000ppm
Operating Principle: 4-electrode electrochemical
Measurement Range: 0-1000 ppm NO
Maximum Overload: 0-100,000ppm
Inboard Filter: To remove acid gases
Filter Life: > 400,000 ppm hours (1000ppm NO @ 200ml/min)
Expected Operating Life: Three years in air
Output Signal: 0.010 ± 0.004 μA/ppm
Resolution: 10ppm
Temperature Range: -20°C to +50°C
Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.007 ± 0.003 %signal/mBar
T90 Response Time: <40 seconds
Relative Humidity Range: 15 to 90% non-condensing
Typical Baseline Range (pure air): -30 to +100ppm equivalent
Maximum Zero Shift (+20°C to +40°C): -100ppm equivalent
Long Term Output Drift: <2% signal loss/month
Recommended Load Resistor: 10Ω
Bias Voltage: Not required
Repeatability: 1% of signal
Output Linearity: Linear
Colour Coding: Red

PHYSICAL CHARACTERISTICS
Weight: 12.5 g nominal
Position Sensitivity: None
Storage Life: Six months in CTL container
Recommended Storage Temp: 0°C to 20°C

Life expectancy
> 3 years
Measurement range
0-40.000 ppm
Target Gas
Carbon Monoxide
Response Time
< 40 seconds (T90)
Output
10 ± 4 nA/ppm
Datasheet47.56 KB
Contact us

Vuoi una quotazione o hai bisogno di informazioni?

I have read and accepted the information relating to the Processing of Personal Data pursuant to EU Regulation 2016/679 articles 13 and 14 * Privacy information