MEASUREMENT
Operating Principle: 3-electrode electrochemical
Measurement Range: 0-2000 ppm CO
Maximum Overload: 4000 ppm CO
Filter: Double Size Filter To remove acid gases
Sensitivity: 0.075 ± 0.025 μA/ppm
Response Time (T90): < 38 seconds
Baseline Offset (clean air): -2 to +17 ppm equivalent
Resolution (when using recommended circuitry): 1 ppm
Repeatability: <1% of signal
Linearity: Linear
ELECTRICAL
Recommended Load Resistor: 10 Ω
Bias Voltage: 0 or +250 mV
MECHANICAL
Weight: 31 g
Housing Material: Polycarbonate
Orientation: Any
ENVIRONMENTAL
Typical Applications: Ambient Environmental Monitoring
Operating Temperature Range: -20°C to +50°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.02% signal/mBar
Operating Humidity Range: 15 - 90% RH non-condensing
LIFETIME
Long Term Sensitivity Drift: <2% signal loss/year
Expected Operating Life: Two years in air
Storage Life: 6 months in CTL container
A3ED
Product code
A3ED
Manufacturer
Life expectancy
> 24 months
Measurement range
0-2000 ppm
Target Gas
Carbon Monoxide
Response Time
< 38 seconds (T90)
Output
75 ± 25 nA/ppm
Datasheet309.37 KB