Contact us
COCl2 3E 1
Product code
COCl2 3E 1
Manufacturer
Measuring Range: 0–1 ppm
Sensitivity Range: 650 nA/ppm ± 150 nA/ppm
Zero Current at 20°C: < ± 10 nA
Resolution at 20°C: < 0.02 ppm
Bias Potential: 0 mV
Linearity: < 10% full scale
Response Time at 20°C:
- t50 < 60 s calculated from 2 min. exposure time at 40% r.H.
- t90 < 120 s calculated from 2 min. exposure time at 40% r.H.
Long Term Sensitivity Drift: < 5% per 6 months
Operation Conditions
Temperature Range: -20°C to +40°C
Humidity Range: 15–90% r.H., non–condensing
Effect of Humidity: no effect
Sensor Life Expectancy: > 12 months
Warranty: 7 months
Life expectancy
> 12 months
Measurement range
0-1 ppm
Target Gas
Phosgene
Response Time
< 120 seconds (T90)
Output
650 ± 150 nA/ppm
Datasheet203.51 KB