GS+7CL2

GS+7CL2
GS+7CL2
Product code
GS+7CL2
Manufacturer
Electrochemical cell for chlorine, ideal for fixed gas detectors.

Measurement Range 0 – 20 ppm
Output Signal 1000 ± 300 nA / ppm
Linearity:  Linear 
Repeatability < ±2% Cl2 equivalent
Response time, T90  <60 s
Maximum Overload 250 ppm
Long-term Output Drift <20% per annum
Recommended Load Resistor 10 ohms
Warranty 1 year from the date of dispatch
Resolution (Electronics dependent) 0.1 ppm typical

Life expectancy
> 12 months
Measurement range
0 - 20 ppm
Target Gas
Chlorine
Response Time
< 60 seconds (T90)
Output
1.0 ± 0.3 μA/ppm
Datasheet1.29 MB