7HLM

7HLM
7HLM
Product code
7HLM
Manufacturer
City Technology - Electrochemical Cell for Hydrogen Sulfide

MEASUREMENT
Operating Principle
: 3-electrode electrochemical
Measurement Range: 0-200 ppm H2S
Maximum overload: 1000 ppm H2S
Filter: None
Sensitivity: 0.37 ± 0.07 μA/ppm
Response Time (T90): <35 Seconds at 20°C
Zero Shift (+20°C to +40°C): < 2 ppm H2S equivalent
Resolution: 0.25 ppm (when used with recommended electronics)
Baseline Offset (clean air): -0.6 to +1.9 ppm H2S equivalent
Repeatability: 2% of signal
Linearity: Linear

ELECTRICAL
Recommended Load Resistor
: 10 Ω
Bias Voltage: Not required

MECHANICAL
Weight
: 12 g
Housing Material:

  • Cap: Polycarbonate
  • Body: ABS

Orientation: Any

ENVIRONMENTAL
Operating Temperature Range: -40°C to +50°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.008 ± 0.002 % signal/mBar
Operating Humidity Range: 15 - 90% RH non-condensing

LIFETIME
Long Term Sensitivity Drift: 
<2% signal loss/month
Expected Operating Life: Two years in air
Storage Life: 6 months in CTL container

Life expectancy
> 24 months
Measurement range
0-200 ppm
Target Gas
Hydrogen sulfide
Response Time
< 35 seconds (T90)
Output
370 ± 70 nA/ppm
Datasheet396.8 KB
Contact us

Vuoi una quotazione o hai bisogno di informazioni?

I have read and accepted the information relating to the Processing of Personal Data pursuant to EU Regulation 2016/679 articles 13 and 14 * Privacy information