3ETO

3ETO
3ETO
Product code
3ETO
Manufacturer
City Technology - Electrochemical Cell for Ethylene Oxide

MEASUREMENT
Operating Principle
: 3-electrode electrochemical
Measurement Range: 0-20 ppm C2H4O
Maximum Overload: 100 ppm C2H4O
Filter: None
Sensitivity: 2.75 ± 0.5 mA/ppm
Resolution: 0.1 ppm
Response Time (T90): < 140 seconds
Baseline Offset (clean air): 0 to +1 ppm equivalent
Zero Shift (+20°C to +40°C): 2 ppm equivalent
Repeatability: 1% of signal

ELECTRICAL
Recommended Load Resistor
: 10 Ω
Bias Voltage: +300 mV

MECHANICAL
Weight
: 22 g
Orientation: Any

ENVIRONMENTAL
Operating Temperature Range: -20°C to +50°C
Recommended Storage Temp: 0°C to 20°C
Operating Pressure Range: Atmospheric ± 10%
Operating Humidity Range: 15 - 90% RH non-condensing

LIFETIME
Long Term Sensitivity Drift
: <5% signal loss/year
Expected Operating Life: Two years in air
Storage Life: 6 months in CTL container

Life expectancy
> 24 months
Measurement range
0-20 ppm
Target Gas
Ethylene Oxide
Response Time
< 140 seconds (T90)
Output
2.75 ± 0.5 μA/ppm
Datasheet295.19 KB
Contact us

Vuoi una quotazione o hai bisogno di informazioni?

I have read and accepted the information relating to the Processing of Personal Data pursuant to EU Regulation 2016/679 articles 13 and 14 * Privacy information